Internal reference: (b)(4).This case was reviewed and investigated according to the manufacturer¿s policy.Block b3: correction: the initial report submitted 11/12/2020 had an inadvertent entry error of 10/30/2020.Correct date is 10/29/2020.Block h3: the device was visually and microscopically inspected, damage was observed.A portion of the distal floppy tip was separated and not returned for investigation.A remaining portion of the distal floppy tip was damaged but still intact to the device.The damaged distal floppy tip was stretched and slightly flared out, presenting an irregular edge; the material is malleable.The scanner body (transducer) was intact and still attached to the device.The probable cause of the reported tip separation is damage in use, likely from an excessive pulling force, as evidenced by the stretched material observed at the distal end of the damaged floppy tip.Block h10: the manufacturing documentation for this device was reviewed and the device met all quality and manufacturing release criteria.Submission of this report does not, in itself, represent a conclusion by the manufacturer and/or authorized representative or the national competent authority that the content of this report is complete or accurate, that the medical device(s) listed failed in any manner and/or that the medical device(s) caused or contributed to the alleged death or deterioration in the state of the health of any person.
|