Model Number 10-88-60 |
Device Problem
Increased Pump Speed (1501)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 10/15/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
There was no patient involvement.Livanova (b)(4) manufactures the s5 mast roller pump.The incident occurred in (b)(6).A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.Livanova initiated an investigation.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|
|
Event Description
|
Livanova (b)(4) received a report that a s5 mast roller pump was giving an error message related to increased pump speed during procedure.There was no report of patient injury.
|
|
Event Description
|
See initial report.
|
|
Manufacturer Narrative
|
H.10: the unit was returned to manufacturer site for repair.The pump head was found blocked and could not be rotated.Dirty and traces of blood could be found on the device.The pump was cleaned and disinfected.The processor board and flat ribbon cable were replaced.Other components were replaced as well due to traces of use.Subsequent functional verification testing was completed without further issues and the unit was returned to service.The most likely root cause of the reported issue is a defective processor board/flat ribbon cable.
|
|