• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COLOPLAST A/S GENESIS MALLEABLE; PENILE PROSTHESIS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COLOPLAST A/S GENESIS MALLEABLE; PENILE PROSTHESIS Back to Search Results
Model Number 5192601400
Device Problems Inadequacy of Device Shape and/or Size (1583); No Apparent Adverse Event (3189)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
The lot # was reviewed for complaint trend, nonconforming report and capa.Devices met specification prior to release and no trends were noted.Without the benefit of examination and testing, coloplast is precluded from commenting on the condition of the device or the cause of the occurrence.Once our evaluation is complete, a follow-up report will be submitted.
 
Event Description
According to the available information, the rods were unequal in size, which caused issues with usage/penetration.The device was removed and replaced.
 
Event Description
This follow-up mdr is created to document the additional event information received for record #622457.According to the available information the malleable penile prosthesis was implanted on (b)(6) 2019 and removed/replaced on (b)(6) 2020 due to usage issues.Additional information received from the territory manager indicated: ¿sized being unequal caused penetration issues.Device was fully intact.¿ further information received indicated: ¿implant not even¿.Two malleable rods were received for evaluation.As examination of the device may not conclusively confirm or disprove the report of use error quality accepts the physician¿s observations as to the reason for surgical intervention.A review of the device history record confirmed the devices from this lot met all specifications prior to release.A review of the complaint history database, nonconformances and capas revealed no trends for this lot.
 
Manufacturer Narrative
Based on the available information, the reported complaint is identified in the risk management documentation and reviewed routinely with management to monitor complaint trends as part of post market surveillance.No corrective action is required at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GENESIS MALLEABLE
Type of Device
PENILE PROSTHESIS
Manufacturer (Section D)
COLOPLAST A/S
holtedam 1
humlebæk 3050
DA  3050
MDR Report Key10836972
MDR Text Key216153605
Report Number2125050-2020-01093
Device Sequence Number1
Product Code FAE
UDI-Device Identifier05708932324748
UDI-Public05708932324748
Combination Product (y/n)N
PMA/PMN Number
K040959
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 01/26/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number5192601400
Device Catalogue Number519260
Device Lot Number6398730
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/23/2020
Initial Date Manufacturer Received 10/19/2020
Initial Date FDA Received11/13/2020
Supplement Dates Manufacturer Received10/19/2020
Supplement Dates FDA Received01/25/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age63 YR
-
-