• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. GII SPC PRI SPACER BLOCK; PRSTHSIS, KNEE, FEMRTIBIL, CMNTD, MTL/PLYMR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. GII SPC PRI SPACER BLOCK; PRSTHSIS, KNEE, FEMRTIBIL, CMNTD, MTL/PLYMR Back to Search Results
Model Number 71441265
Device Problem No Apparent Adverse Event (3189)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/26/2020
Event Type  malfunction  
Manufacturer Narrative
Result of investigation: the associated device, used in treatment, was returned and evaluated.A visual inspection of the returned device confirmed the stated failure mode.The white rings on the device are fractured and missing, rendering the device inoperable.The device was manufactured in 2014 and shows signs of extensive use.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary. we consider this investigation closed.
 
Event Description
It was reported, that during tka the white o rings are gone and will not hold on clip on build ups.The procedure finished with the same device.No surgical delay.No patient harm.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GII SPC PRI SPACER BLOCK
Type of Device
PRSTHSIS, KNEE, FEMRTIBIL, CMNTD, MTL/PLYMR
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
1450 brooks rd.
memphis, TN 38116
MDR Report Key10837290
MDR Text Key216174589
Report Number1020279-2020-06429
Device Sequence Number1
Product Code HRY
UDI-Device Identifier00885556034354
UDI-Public00885556034354
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K121393
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Reporter Occupation Other
Type of Report Initial
Report Date 11/13/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model Number71441265
Device Catalogue Number71441265
Device Lot Number14AM01009
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/22/2020
Initial Date Manufacturer Received 10/28/2020
Initial Date FDA Received11/13/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Type of Device Usage N
Patient Sequence Number1
-
-