MAUDE Adverse Event Report: VYAIRE MEDICAL MEDISORB MULTI-ABSORBER ORIGINAL, DISPOSABLE; ABSORBENT, CARBON-DIOXIDE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

VYAIRE MEDICAL MEDISORB MULTI-ABSORBER ORIGINAL, DISPOSABLE; ABSORBENT, CARBON-DIOXIDE Back to Search Results
Model Number MEDISORB¿ MULTI-ABSORBER ORIGINAL, DISPOSABLE
Device Problem Crack (1135)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/19/2020
Event Type  malfunction  
Manufacturer Narrative
At this time, the suspect device has not been returned for evaluation.Therefore, no root cause could be determined yet.Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.
 
Event Description
The customer reported that the medisorb¿ multi-absorber canister has a crack at the bottom, which lead to leakage.The event occurred while connected to a patient however, the customer confirmed that no patient harm was associated on this event.
 
Manufacturer Narrative
The suspect device was not returned for evaluation.However, device history record review reveals no deviations to specification of device.All units packed passed leak test and visual inspection.Batch documentation review reveals no issues that could be linked to the reported failure.Also, devices are packaged in a carton that meets the vyaire drop testing specification and labeled as fragile and to handle with care.Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDISORB MULTI-ABSORBER ORIGINAL, DISPOSABLE
Type of Device
ABSORBENT, CARBON-DIOXIDE
Manufacturer (Section D)
VYAIRE MEDICAL
26125 n. riverwoods blvd.
mettawa IL 60045
MDR Report Key10841248
MDR Text Key216894474
Report Number3002807637-2020-00008
Device Sequence Number1
Product Code CBL
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign
Type of Report Initial,Followup
Report Date 10/19/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMEDISORB¿ MULTI-ABSORBER ORIGINAL, DISPOSABLE
Device Catalogue Number8003138
Device Lot Number3900520, 0610120, 2190320
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 10/19/2020
Initial Date FDA Received11/14/2020
Supplement Dates Manufacturer Received01/04/2021
Supplement Dates FDA Received01/20/2021
Is This a Reprocessed and Reused Single-Use Device? Yes
Patient Sequence Number1
-
-