SMITH & NEPHEW, INC. FLEXIBLE SHAFT W/CIR CONNECTOR; PROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED Back to Search Results
Model Number 71118200
Device Problem Break (1069)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/28/2020
Event Type  malfunction  
Event Description
It was reported that, after procedure, it was noticed that the tip of the flexible shaft w/cir connector has gone off where the reamer heads connect.No case involved.
 
Manufacturer Narrative
The associated device, intended for use in treatment, was returned and evaluated.A visual inspection of the returned device confirmed the stated failure mode.The shaft is fractured, rendering the device inoperable.The device was manufactured in 2012 and shows signs of extensive use.A review of complaint history on the listed part revealed no prior complaint for the listed batch with the same failure mode.A review of the manufacturing records did not reveal any manufacturing or material abnormalities that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLEXIBLE SHAFT W/CIR CONNECTOR
Type of Device
PROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10843191
MDR Text Key216340704
Report Number1020279-2020-06469
Device Sequence Number1
Product Code HSD
UDI-Device Identifier03596010436528
UDI-Public03596010436528
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 03/29/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71118200
Device Catalogue Number71118200
Device Lot Number12GMO3312
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 10/28/2020
Initial Date FDA Received11/16/2020
Supplement Dates Manufacturer Received03/17/2021
Supplement Dates FDA Received03/29/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-