BOSTON SCIENTIFIC CORPORATION WALLFLEX DUODENAL; STENT,METALLIC,EXPANDABLE,DUODENAL Back to Search Results
Model Number M00565020
Device Problem Positioning Problem (3009)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/16/2020
Event Type  Injury  
Manufacturer Narrative
(b)(4).The complainant indicated that the device remains implanted and will not be returned for evaluation; therefore, a failure analysis of the complaint device could not be completed.If any further relevant information is identified, a supplemental medwatch will be filed.
 
Event Description
It was reported to boston scientific corporation on (b)(6) 2020 that a wallflex duodenal stent was implanted to treat a 4cm duodenal stricture during an esophagogastroduodenoscopy (egd) with stent placement procedure performed on (b)(6) 2020.Reportedly, the patient's anatomy was tortuous and was dilated prior to stent placement.According to the complainant, during the procedure, the stent was deployed.However, it was observed under a video monitor that the stent moved out from the target location.The physician attempted to use a rat tooth forceps to reposition the stent; however, it was unsuccessful.The stent remains implanted and another wallflex duodenal stent was implanted to complete the procedure.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WALLFLEX DUODENAL
Type of Device
STENT,METALLIC,EXPANDABLE,DUODENAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key10844381
MDR Text Key216393545
Report Number3005099803-2020-05354
Device Sequence Number1
Product Code MUM
UDI-Device Identifier08714729456490
UDI-Public08714729456490
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K062750
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 11/16/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/29/2022
Device Model NumberM00565020
Device Catalogue Number6502
Device Lot Number0026095475
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/24/2020
Initial Date FDA Received11/16/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/29/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age59 YR
-
-