Qn#(b)(4).The customer reported that the oxygen tubing disconnected from the nebulizer at 6-8 lpm.The customer returned components from catalog number 1734 up-draft ii opti-neb nebulizer w/reserv, batch 74b2002407.During the visual inspection, it was noted that a jet, jar, cap, tubing, mouthpiece, tee-connector and corrugated tubing were returned for analysis.During the visual inspection, no defects or anomalies were found with the returned components.A functional inspection was performed to determine if the oxygen tubing disconnects from the nebulizer at 6-8 lpm.The returned tubing was used to connect the nebulizer unit to the air flowmeter.5cc of water was added to the returned nebulizer unit and tubing was connected to an air flowmeter and the pressure was increased to 8 lpm.During functional testing, mist was produced from the chamber of the nebulizer and the oxygen tubing did not disconnect from the nebulizer.No problems were found with the returned sample.Based on the investigation performed, the reported complaint could not be confirmed.There were no issues found with the returned device.The unit functioned as intended.
|