SMITH & NEPHEW, INC. FEMORAL HEAD/NECK IMPACTOR; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/CERAMIC/POLYMER, CEMENTED OR NON-POROUS Back to Search Results
Model Number 71360093
Device Problem Defective Device (2588)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/25/2020
Event Type  Injury  
Event Description
It was reported that during thr procedure the plastic of the femoral head/neck impactor got damaged.This happened during use outside the patient.There was no reported delay.Procedure concluded with a competitor device.No injuries or other complications were reported at this time.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Manufacturer Narrative
Lot number and expiration date added the device, used in treatment was returned for evaluation.Visual inspection of the returned device confirms that the device is damaged and has signs of wear and tear from use.According to clinical/medical investigation, this case reports the plastic component of the femoral head pusher was damaged during surgery.Per complaint detail, there was no patient injury or surgical delay, and the procedure was completed using a competitor device.Since no patient harm is alleged, no further clinical assessment is warranted.A review of the manufacturing records did not reveal any manufacturing or material abnormalities that could have caused or contributed to the reported incident.A review of complaint history for the listed part revealed no prior complaints for the lot.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.The device is a reusable device that can be exposed to numerous surgeries and cleaning cycles.Probable causes that could contribute to the reported event, as reusable instruments are susceptible to damage from prolonged use and through misuse or rough handling.To avoid compromised performance or damage, it is recommended the device be inspected prior to and after each use and cleaning.We recommend that all re-usable instruments be routinely inspected for wear/damage and replaced as necessary.A review of risk management files and the instructions for use found that the reported failure was documented appropriately.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further actions are being taken at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FEMORAL HEAD/NECK IMPACTOR
Type of Device
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/CERAMIC/POLYMER, CEMENTED OR NON-POROUS
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10844435
MDR Text Key216388581
Report Number1020279-2020-06481
Device Sequence Number1
Product Code LZO
UDI-Device Identifier03596010619310
UDI-Public03596010619310
Combination Product (y/n)N
PMA/PMN Number
K113789
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 02/08/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71360093
Device Catalogue Number71360093
Device Lot Number20AM06938
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 10/25/2020
Initial Date FDA Received11/16/2020
Supplement Dates Manufacturer Received02/04/2021
Supplement Dates FDA Received02/08/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-