Model Number CV-180
Device Problem Loss of Power (1475)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
The suspect device was returned to olympus and evaluated.The reported problem could not be confirmed.The unit was burned in for more than 3 hours with a test clv-180 and the customer's clv-180; the power functioned as expected and did not shut off during testing.However, corrosion and debris was found in the video connector and it was determined replacement of the video connector is necessary.The device was repaired and verified to operate within specifications and returned to the customer.
 
Event Description
A user facility reported to olympus that the device "got wet, was not working properly and shut off." the problem, as reported to olympus, occurred during a procedure.There was no patient injury or harm, associated with the problem, reported to olympus.
 
Manufacturer Narrative
This supplemental report is submitted to provide the results of the legal manufacturer¿s investigation and device history record (dhr) review.The dhr for the subject device was reviewed and it was verified the device was manufactured in accordance with documented specifications.The legal manufacturer performed an investigation.A conclusive root cause was not identified.The legal manufacturer determined the likely cause of the event was failure of the power supply unit due to water invasion, occurring when the user spilled water on the unit.The cause of the corrosion and debris in the video connector, found on the device evaluation, was likely due to user handling.As stated in the instructions for use, as a preventive measure, "after wiping with a piece of moistened gauze, dry the video system center thoroughly before using it again.If it is used while still wet, there is the risk of an electric shock.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA II VIDEO SYSTEM CENTER
Type of Device
VIDEO SYSTEM CENTER
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
MDR Report Key10855935
MDR Text Key225931126
Report Number8010047-2020-09027
Device Sequence Number1
Product Code FAJ
UDI-Device Identifier04953170215513
UDI-Public04953170215513
Combination Product (y/n)N
PMA/PMN Number
K133538
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Type of Report Initial,Followup
Report Date 01/14/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCV-180
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/26/2020
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 10/20/2020
Initial Date FDA Received11/17/2020
Supplement Dates Manufacturer Received12/18/2020
Supplement Dates FDA Received01/14/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-