It was reported that, during tka surgery, profix femoral impactor broke.All fragments were removed from the patient.Procedure was concluded with the same device, without a delay or an injury to the patient.
|
The device, used in treatment, was returned for evaluation.The clinical/medical team concluded, this case reports femoral impactor broke while being used outside of the patient.Per email communication, no broken pieces had to be removed from the patient and the procedure was completed using the same device, without a surgical delay or injury to the patient.Therefore, since no patient harm is alleged, no further assessment is warranted at this time.A visual inspection confirmed the tip is broken off the profix oxinium fem impactor.The device shows significant signs of wear/usage.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we have no reason to suspect that the products failed to meet any product specifications at the time of manufacture.A review of complaint history for the listed part revealed no prior complaints for the listed batch a review of the risk management file revealed this failure mode was previously identified.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.This is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.Case-2020-00025260-1.
|