The subject device was returned to omsc for evaluation and checked the subject device.There were no scratches, holes, or clogging with the foreign objects inside the subject device, but it was found the dent at 5cm instrument channel of the bending section from the distal end of the subject device.Also it was test with running inside the subject device with the cleaning brushes, bw-412t which is same model of user use, and bw-20t.The brushes run smoothly.The dent of the instrument channel of the bending section was surmised that might have occurred due to excessive bending stress apply to the bending section of the subject device, not the endotherapy instruments.Omsc reviewed the manufacturing history (dhr) of the subject device and confirmed no irregularity.The exact cause has been under investigation, therefore the exact cause of the reported event could not be conclusively determined at this time.If additional information becomes available, this report will be supplemented.
|