• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER SURGICAL, INC. 7MM FULL PERF FLAT DRAIN 7MM 100% SI; APPARATUS, SUCTION, SINGLE PATIENT USE, PORTABLE, NONPOWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER SURGICAL, INC. 7MM FULL PERF FLAT DRAIN 7MM 100% SI; APPARATUS, SUCTION, SINGLE PATIENT USE, PORTABLE, NONPOWERED Back to Search Results
Model Number N/A
Device Problem Detachment of Device or Device Component (2907)
Patient Problem Foreign Body In Patient (2687)
Event Date 10/01/2020
Event Type  Injury  
Manufacturer Narrative
This event has been recorded by zimmer biomet under (b)(4).Once the investigation is complete, a follow up/final report will be submitted.Not available for evaluation.
 
Event Description
It was reported that the patient underwent l5 to s1 lumbar spine fusion surgery and a lumbar drain placed.The drain was being removed on post-operative day four when it fractured.The patient returned to the or for removal of the fractured piece.No additional consequences have been reported as a result of this malfunction.
 
Event Description
No additional event information available.
 
Manufacturer Narrative
This event has been recorded by zimmer biomet under (b)(4).This medwatch is being filed to relay additional information.The following sections were updated/corrected: b4, b5, g4, g7, h2, h3, h6, h8, h10.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.Lot identification is necessary for review of device history records, and lot identification was not provided.Device is used for treatment.Medical records were not provided.A definitive root cause cannot be determined.The event cannot be confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.H3 other text : device not returned.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
7MM FULL PERF FLAT DRAIN 7MM 100% SI
Type of Device
APPARATUS, SUCTION, SINGLE PATIENT USE, PORTABLE, NONPOWERED
Manufacturer (Section D)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
MDR Report Key10867986
MDR Text Key217102029
Report Number0001526350-2020-00981
Device Sequence Number1
Product Code GCY
Combination Product (y/n)N
PMA/PMN Number
PREAMENDMENT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,other,use
Type of Report Initial,Followup
Report Date 01/26/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number00256900010
Device Lot NumberN/A
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 10/28/2020
Initial Date FDA Received11/19/2020
Supplement Dates Manufacturer Received01/22/2021
Supplement Dates FDA Received01/26/2021
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Outcome(s) Other; Required Intervention;
Patient Age53 YR
Patient Weight59
-
-