Model Number N/A |
Device Problem
Device Contaminated During Manufacture or Shipping (2969)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 10/28/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Report source (b)(6).Product is in process of being returned to zimmer biomet for the investigation and a follow-up mdr will be submitted upon completion.Multiple mdr reports were filed for this event, please see associated reports: 0001825034 - 2020 - 04146.
|
|
Event Description
|
It was reported that during incoming inspection, debris was found in the sterile packaging.There was no patient involvement.Attempts have been made and additional information on the reported event is unavailable at this time.
|
|
Manufacturer Narrative
|
This follow-up report is being submitted to relay additional information.Complaint sample was returned and evaluated against the reported event.Visual inspection of the returned product identified that there is debris inside the sterile package.Dhr was reviewed and no discrepancies were found.The likely condition of the products when they left zimmer biomet was non-conforming.The root cause of the reported event is the operator not following instructions during the manufacturing process.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
No additional event information to report at this time.
|
|