Device evaluated by mfr.: returned product consisted of an angiojet spiroflex catheter.The pump assembly, effluent/supply line, shaft, tip and spike line were visually examined for damage or any irregularities.It was noticed when returned there was 25cc of blood in the waste bag and the device.The shaft showed multiple kinks.Functional testing was competed.The device was inserted into the test ultra drive unit console and the device was tested.The device would not prime due to the damaged shaft.There were 2 leaks along the shaft approximately 110cm and 114cm from the tip.Shaft damage at these locations in the outer shaft was noticed.The pressure was verified and was below the normal operating range.No other failures or error codes were noticed during the testing.Inspection of the remainder of the device, apart from the observed damage revealed no other damage or irregularities.
|