|
Model Number CATRXKIT |
Device Problem
Suction Failure (4039)
|
Patient Problems
No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/28/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
Results: the catrx was kinked approximately 38.0 cm from the hub.Conclusions: evaluation of the returned catrx revealed a kink.If the catrx is forcefully mishandled at extreme angles during use, damage such as a kink may occur.During functional testing, low aspiration was observed from the catrx while attempting to aspirate water through the catheter.The kink likely contributed to the catrx's inability to aspirate during the procedure and low aspiration during the functional test.Penumbra catheters are inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|
|
Event Description
|
The patient was undergoing a thrombectomy procedure in the left anterior descending artery (lad) using an indigo system catrx aspiration catheter (catrx) and aspiration tubing (tubing).During the procedure, the catrx would not aspirate when connected to the tubing.Subsequently, the physician disconnected the catrx from the tubing and discovered that the tubing would aspirate; however, the catrx would not aspirate when connected to the tubing.Therefore, the catrx was removed.The procedure was completed using a new catrx and the same tubing.There was no report of an adverse effect to the patient.
|
|
Search Alerts/Recalls
|
|
|