Philips has investigated this complaint.According to the information collected, the hospital staff noticed a burning smell.The staff identified that the burning smell was coming from the multidiagnost eleva system that was located at the same department.The multidiagnost eleva system was switched off, the fire alarm activated and the department evacuated.There were no signs of smoke or flames.The system was not in clinical use at the time of the event.A philips engineer checked the system onsite and could not identify any burned components.The burning smell came from the tube cooling unit that had been replaced in the morning of the day of the event.The most probable cause was smoldering dust of the sound proofing insulating foam of the cover of the tube cooling unit.The insulating foam was removed.
|