Model Number 5392
Device Problem Output above Specifications (1432)
Patient Problems Shaking/Tremors (2515); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/19/2020
Event Type  Injury  
Manufacturer Narrative
Product analysis: analysis was unable to confirm the customer comment of excessive current from the epg.No fault was found.The epg passed the visual check and booted up normally.The device was returned to the customer unrepaired.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that an excessive current from the external pulse generator (epg) caused the patient to "tremble".The epg was replaced with another during the procedure to complete the case.The epg was returned for repair.No further patient complications have been reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DUAL CHAMBER TEMPORARY PACEMAKER
Type of Device
PULSE-GENERATOR, PACEMAKER, EXTERNAL
Manufacturer (Section D)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer (Section G)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key10873803
MDR Text Key217287123
Report Number2182208-2020-02903
Device Sequence Number1
Product Code DTE
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K150246
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 11/20/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number5392
Device Catalogue Number5392
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/26/2020
Initial Date Manufacturer Received 10/26/2020
Initial Date FDA Received11/20/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/16/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-