• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US ATTUNE PS FEM TRIAL SZ 6 RT; ATTUNE INSTRUMENTS : FEMORAL TRIALS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US ATTUNE PS FEM TRIAL SZ 6 RT; ATTUNE INSTRUMENTS : FEMORAL TRIALS Back to Search Results
Catalog Number 254500726
Device Problem Break (1069)
Patient Problems No Consequences Or Impact To Patient (2199); No Code Available (3191)
Event Date 11/09/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Trial spring was damaged and femur trial was expired or ruptured.Products: fb insert test sz 6 and femoral test sz 6 ps.
 
Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event Description
Additional information received indicated that the femoral test was cracked or broken and the inserted test spring was damaged during cleaning.There were no intraoperative complications or increased surgical time.
 
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : the device associated with this report was not returned for evaluation.Visual examination of the provided photograph found that the instrument was cracked.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot : the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).Where the lot code was provided, a manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE PS FEM TRIAL SZ 6 RT
Type of Device
ATTUNE INSTRUMENTS : FEMORAL TRIALS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key10875209
MDR Text Key219563619
Report Number1818910-2020-25098
Device Sequence Number1
Product Code HWT
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup,Followup,Followup
Report Date 11/09/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number254500726
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 11/09/2020
Initial Date FDA Received11/20/2020
Supplement Dates Manufacturer Received11/09/2020
01/18/2021
01/21/2021
Supplement Dates FDA Received12/05/2020
01/19/2021
01/22/2021
Is This a Reprocessed and Reused Single-Use Device? No