Upon receipt, the lead under complaint was subjected to an extensive analysis.The performance of the device was scrutinized, including a visual and electrical inspection.The visual inspection revealed 32 cm distal to the is4 connector pin that the lead body was found squeezed and deformed.In this area the conductor coils leading to the ring electrodes and to the lead tip were found fractured, which is assumed to be the root cause of the reported clinical observation.Based on the characteristics, as well as the location of the damage, it is reasonable to assume that the lead was subject to excessive mechanical forces as the result of clavicular first rib entrapment.The manufacturing process for this device was re-investigated.All production steps had been performed accordingly.There was no sign of any inconsistency during the manufacturing process.In conclusion, the analysis did not reveal any sign of a material or manufacturing problem.
|