Model Number 71440890
Device Problem Break (1069)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/28/2020
Event Type  malfunction  
Event Description
It was reported that during a tka procedure, the gii oxinium femoral impactor broke.It is unknown how the surgery was completed and if there was a delay.The patient outcome is unknown as well.
 
Manufacturer Narrative
H6,h10: the associated device, used in treatment, was returned and evaluated.A visual inspection of the returned device confirmed the stated failure mode.The bumper on the device is fractured, rendering the device inoperable.The device was manufactured in 2010 and shows signs of extensive use.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history did not reveal additional complaints for the listed batch.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GII OXINIUM FEMORAL IMPACTOR
Type of Device
PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10878855
MDR Text Key217647213
Report Number1020279-2020-06605
Device Sequence Number1
Product Code JDG
UDI-Device Identifier03596010476074
UDI-Public03596010476074
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup
Report Date 12/10/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71440890
Device Catalogue Number71440890
Device Lot Number10CM01360
Initial Date Manufacturer Received 10/28/2020
Initial Date FDA Received11/20/2020
Supplement Dates Manufacturer Received12/03/2020
Supplement Dates FDA Received12/10/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-