SMITH & NEPHEW, INC. TANDEM STR THREADED HNDLE; PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL
|
Back to Search Results |
|
Model Number 71362804 |
Device Problem
Break (1069)
|
Patient Problems
No Known Impact Or Consequence To Patient (2692); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/26/2020 |
Event Type
malfunction
|
Event Description
|
It was reported that during efip inspection, both instruments cracked at the handles and easily broke apart while handling.No case involved.
|
|
Manufacturer Narrative
|
The devices, intended for use in treatment were returned for evaluation: a visual inspection was performed and confirmed the two device¿s handles are broken off.The manufactured date for this device is 2008.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A complaint history review found related failures; this failure mode will be monitored for future complaints for any necessary corrective actions.These devices are reusable instruments that can be exposed to numerous surgeries; damage from repeated use can occur.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.Should additional information be received, the complaint will be reopened.We consider this investigation closed.
|
|
Search Alerts/Recalls
|
|
|