• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DENTSPLY LLC 30K FOCUSED SPRAY SLIMLINE ULTRASONIC INSERT - 1000 (LOW-HIGH POWER); SCALER, ULTRASONIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DENTSPLY LLC 30K FOCUSED SPRAY SLIMLINE ULTRASONIC INSERT - 1000 (LOW-HIGH POWER); SCALER, ULTRASONIC Back to Search Results
Catalog Number 81570
Device Problems Restricted Flow rate (1248); Overheating of Device (1437); Infusion or Flow Problem (2964)
Patient Problem Insufficient Information (4580)
Event Type  malfunction  
Manufacturer Narrative
There has been a previous report received where lack of water flow has caused an overheating insert.Since an overheating insert could necessitate medical/surgical intervention to preclude permanent damage to a body structure or permanent impairment of a body function, this malfunction would be likely to cause/contribute to a serious injury should it recur.As such, this event meets the criteria for reportability per 21 cfr part 803.The device is available for evaluation, though has not been returned as of this report.Evaluation results will be submitted as they become available.
 
Event Description
While using a 30k fsi-sli-1000 insert, there was a stack or water flow issue producing extra heat within the sterimate.The event outcome is unknown as of this mdr evaluation.
 
Manufacturer Narrative
Visually inspected and verified the insert has water and vibration meets spec.The insert was tested on a digital thermometer gauge # 6116 due date: (b)(6) 2021.The temperature is 84.2° degrees meets spec.Specification states not to exceed 118.4° f.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
30K FOCUSED SPRAY SLIMLINE ULTRASONIC INSERT - 1000 (LOW-HIGH POWER)
Type of Device
SCALER, ULTRASONIC
Manufacturer (Section D)
DENTSPLY LLC
1301 smile way
york PA 17404
MDR Report Key10881542
MDR Text Key218679768
Report Number2424472-2020-00071
Device Sequence Number1
Product Code ELC
Combination Product (y/n)N
PMA/PMN Number
K052334
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Type of Report Initial,Followup
Report Date 02/15/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number81570
Device Lot Number00044471
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received11/21/2020
Supplement Dates Manufacturer Received02/15/2021
Supplement Dates FDA Received02/15/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-