Model Number 81000 |
Device Problems
Mechanical Problem (1384); Device Slipped (1584); Unintended Movement (3026)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/31/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
Investigation: a terumo bct service technician checked out the machine at the customer site.The technician deternined that the device needed adjustment/realignment and was able to duplicate the reported condition.The technician mechanically adjusted the iv pole stop as needed and verified proper operation.The device serial number history report indicates no further related issues have been reported for this device.One year of service history was reviewed for this device with no other problems identified related to the reported condition.Investigation is in process.A follow-up report will be provided.
|
|
Event Description
|
The customer reported that the iv pole on the trima equipment unexpectedly lowers down.No injury was reported for this incident and no patient was connected at the time the iv pole was sliding down, therefore no patient information is reasonably known.
|
|
Manufacturer Narrative
|
This report is being filed to provide additional information in g.4, h.4, h.6 and h.10.Corrected information is provided in d.2.Root cause: the root cause of this failure was the adjustment of the iv pole.
|
|