Model Number 81000
Device Problems Mechanical Problem (1384); Device Slipped (1584); Unintended Movement (3026)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/31/2020
Event Type  malfunction  
Manufacturer Narrative
Investigation: a terumo bct service technician checked out the machine at the customer site.The technician deternined that the device needed adjustment/realignment and was able to duplicate the reported condition.The technician mechanically adjusted the iv pole stop as needed and verified proper operation.The device serial number history report indicates no further related issues have been reported for this device.One year of service history was reviewed for this device with no other problems identified related to the reported condition.Investigation is in process.A follow-up report will be provided.
 
Event Description
The customer reported that the iv pole on the trima equipment unexpectedly lowers down.No injury was reported for this incident and no patient was connected at the time the iv pole was sliding down, therefore no patient information is reasonably known.
 
Manufacturer Narrative
This report is being filed to provide additional information in g.4, h.4, h.6 and h.10.Corrected information is provided in d.2.Root cause: the root cause of this failure was the adjustment of the iv pole.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIMA ACCEL
Type of Device
TRIMA ACCEL AUTOMATED BLOOD COLLECTION SYSTEM
Manufacturer (Section D)
TERUMO BCT
lakewood CO 80215
MDR Report Key10886156
MDR Text Key220980363
Report Number1722028-2020-00521
Device Sequence Number1
Product Code GKT
UDI-Device Identifier05020583810006
UDI-Public05020583810006
Combination Product (y/n)N
PMA/PMN Number
BK180231
Number of Events Reported1
Summary Report (Y/N)Y
Report Source Manufacturer
Source Type health professional
Type of Report Initial,Followup
Report Date 11/23/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number81000
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? Yes
Initial Date Manufacturer Received 10/31/2020
Initial Date FDA Received11/23/2020
Supplement Dates Manufacturer Received02/05/2021
Supplement Dates FDA Received02/17/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
-
-