ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER CENTURION VISION SYSTEM, ACCESSORY, ACTIVE SENTRY HANDPIECE; UNIT, PHACOFRAGMENTATION Back to Search Results
Catalog Number 8065752914
Device Problem Incorrect, Inadequate or Imprecise Result or Readings (1535)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event Description
A nurse reported that during a surgical procedure the handpiece failed, there was no phacoemulsification power.The handpiece was exchanged and the procedure was completed with no harm to the patient.Additional information was requested an received.
 
Manufacturer Narrative
The phacoemulsification (phaco) handpiece was received and a visual assessment of the returned sample found no visual nonconformities.The phaco handpiece was connected to a calibrated resistance breakout box, where the output and input impedances were found to be within specification.The returned phaco handpiece was connected to a calibrated system.The phaco handpiece tuned successfully and completed a five-minute burn-in test with the system set at 100% ultrasonic and torsional power.The phaco handpiece was connected to dynamic tuning fixture (dtf) for stroke length testing on the longitudinal and torsional movements, which found the handpiece to meet product specifications.The phaco handpiece manufacturing device history record (dhr) was reviewed.Based on assessment, the product met specifications at the time of release.The phaco handpiece was found to meet specifications; therefore, the root cause of the reported event cannot be determined conclusively.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTURION VISION SYSTEM, ACCESSORY, ACTIVE SENTRY HANDPIECE
Type of Device
UNIT, PHACOFRAGMENTATION
Manufacturer (Section D)
ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
MDR Report Key10886218
MDR Text Key218907013
Report Number2028159-2020-01002
Device Sequence Number1
Product Code HQC
Combination Product (y/n)N
PMA/PMN Number
K161794
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 03/26/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number8065752914
Device Lot Number13CFU2
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/11/2020
Initial Date Manufacturer Received 11/02/2020
Initial Date FDA Received11/23/2020
Supplement Dates Manufacturer Received03/19/2021
Supplement Dates FDA Received03/26/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-