ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER CENTURION VISION SYSTEM, ACCESSORY, ACTIVE SENTRY HANDPIECE; UNIT, PHACOFRAGMENTATION
|
Back to Search Results |
|
Catalog Number 8065752914 |
Device Problem
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
|
|
Event Description
|
A nurse reported that during a surgical procedure the handpiece failed, there was no phacoemulsification power.The handpiece was exchanged and the procedure was completed with no harm to the patient.Additional information was requested an received.
|
|
Manufacturer Narrative
|
The phacoemulsification (phaco) handpiece was received and a visual assessment of the returned sample found no visual nonconformities.The phaco handpiece was connected to a calibrated resistance breakout box, where the output and input impedances were found to be within specification.The returned phaco handpiece was connected to a calibrated system.The phaco handpiece tuned successfully and completed a five-minute burn-in test with the system set at 100% ultrasonic and torsional power.The phaco handpiece was connected to dynamic tuning fixture (dtf) for stroke length testing on the longitudinal and torsional movements, which found the handpiece to meet product specifications.The phaco handpiece manufacturing device history record (dhr) was reviewed.Based on assessment, the product met specifications at the time of release.The phaco handpiece was found to meet specifications; therefore, the root cause of the reported event cannot be determined conclusively.The manufacturer internal reference number is: (b)(4).
|
|