|
Model Number LCS4K30BTOL |
Device Problem
Defective Device (2588)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 10/28/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
Device was received and evaluated at olympus (b)(4).Visual inspection was performed and actual product was noted to be in a damaged state.Device evaluation confirmed that the stopper at the tip of the sheath was found deformed.The device history records for this product have been reviewed.All records indicate that the product was manufactured, tested in accordance with all applicable procedures, and met all final product release criteria.As the actual product is clearly in a damaged state and is at a level detected by inspection at the time of shipment, it is considered unlikely that it was damaged from the time of shipment.Olympus will continue to monitor complaints for this device.
|
|
Event Description
|
It was reported that during preparation for use for a functional endoscopic sinus surgery (fess) procedure, the user opened the lcs4k30bto device and found that the stopper at the tip of the shaft was found rolled up,deformed.The device was replaced with similar device and the intended procedure was completed.No patient harm or impact was reported.No user injury reported.
|
|
Search Alerts/Recalls
|
|
|