ST. JUDE MEDICAL INQUIRY AFOCUSII (4 F DOUBLE LOOP) CATHETER 20 ELECTRODE, 7F; CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING
|
Back to Search Results |
|
Model Number IBI-87008 |
Device Problems
Material Frayed (1262); Difficult to Remove (1528); Activation, Positioning or Separation Problem (2906)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 11/11/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
The results/method and conclusion codes along with investigation results will be provided in the final report.
|
|
Event Description
|
During the procedure, after the catheter was used in the left atrium and when attempting to perform mapping in the right atrium, the catheter suddenly stopped deflecting.The catheter was not stuck in the valve.The catheter was attempted to be removed from the patient; however, it could not be removed from the 10f sl0 introducer.It could be finally removed after pulling back and forward with force.When the catheter was checked, a part of the blue coating was peeled off and the wires were found to be exposed.The catheter could be deflected after being removed from the patient.The catheter was replaced and the procedure was completed with no adverse consequences to the patient.
|
|
Event Description
|
During the procedure, after the catheter was used in the left atrium and when attempting to perform mapping in the right atrium, the catheter suddenly stopped deflecting.The catheter was not stuck in the valve.The catheter was attempted to be removed from the patient; however, it could not be removed from the 10f sl0 introducer.It could be finally removed after pulling back and forward with force.When the catheter was checked, a part of the blue coating was peeled off and the wires were found to be exposed.The catheter could be deflected after being removed from the patient.The introducer was aspirated before the replacement catheter was inserted to check if there was any fragment left and it was confirmed that no fragment was left in the introducer and the patient under fluoroscopy.The catheter was replaced and the procedure was completed with no adverse consequences to the patient.
|
|
Manufacturer Narrative
|
One duodecapolar, double loop, inquiry afocusii diagnostic catheter was received for evaluation.The reported damage to the catheter shaft and deflection issue was confirmed, however the removal difficulty could not be confirmed.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the catheter damage and deflection issue are consistent with damage during use.The cause of the withdrawal difficulty remains unknown.
|
|
Search Alerts/Recalls
|
|
|