Model Number 2941 |
Device Problem
Patient-Device Incompatibility (2682)
|
Patient Problem
Pressure Sores (2326)
|
Event Date 10/19/2020 |
Event Type
Injury
|
Event Description
|
It was reported a patient developed a bed sore from a stryker mattress.No information has been provided regarding the severity or treatment of the bed sores.Attempts have been made to reach the customer for more information; however, the customer has not responded to these attempts.
|
|
Manufacturer Narrative
|
Section b5 has been updated to reflect additional information provided by the user facility.The user facility was not able to make the device available for evaluation and stated that the pressure injury was a result of use error, section h codes have been updated to reflect this.
|
|
Event Description
|
It was reported a patient developed a bed sore from a stryker mattress.Upon further investigation it was found that the user facility disconnected the air flow to the mattress, which caused the surface to completely deflate.The patient was not able to be moved to a different surface and developed a bed sore.
|
|