|
Model Number CATRXKIT |
Device Problem
Suction Failure (4039)
|
Patient Problems
No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/29/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
Results: no visible damage was observed on the catrx.Coagulated blood was observed within the catrx.Conclusions: evaluation of the returned catrx confirmed that the device was unable to produce aspiration.During functional testing, the catrx was connected to the returned tubing, and the tubing was connected to a demonstration pump and canister.The pump was powered on, and no aspiration was observed.While attempting to advance a test mandrel through the catrx, resistance was encountered, and the test mandrel could not be advanced through the catheter.The root cause of the catrx not aspirating during the procedure and during functional testing could not be determined.Further evaluation revealed coagulated blood within the catheter.This damage was likely incidental to the complaint.The returned tubing was undamaged and was able to aspirate water without issue.Penumbra catheters and tubing are inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.(b)(4).
|
|
Event Description
|
The patient was undergoing a thrombectomy procedure in the right coronary artery (rca) using an indigo system catrx aspiration catheter (catrx) and aspiration tubing (tubing).During the procedure, when connected to the tubing, the catrx would not aspirate.Subsequently, the physician disconnected the tubing, and the tubing was able to aspirate.However, when reconnecting the catrx to the tubing, the catrx would not aspirate.Therefore, the catrx was removed.The procedure was completed using a new catrx and the same tubing.There was no report of an adverse effect to the patient.
|
|
Search Alerts/Recalls
|
|
|