Model Number CD1357-40Q
Device Problems High impedance (1291); Loose or Intermittent Connection (1371); Failure to Sense (1559)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/11/2020
Event Type  Injury  
Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Event Description
It was reported that an asymptomatic patient presented to a follow-up in-clinic on (b)(6) 2020 after receiving a patient notifier for high out of range right ventricular lead pacing impedance and failure to sense.It was suspected to be caused by a loose implantable cardioverter defibrillator set-screw.The patient underwent reoperation on (b)(6) 2020.The loose set screw was confirmed as the cause of the issues and was addressed.Patient was stable after the procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FORTIFY ASSURA VR ICD, US
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
sonali arangil
15900 valley view court
sylmar, CA 91342
MDR Report Key10905799
MDR Text Key218499143
Report Number2017865-2020-19389
Device Sequence Number1
Product Code LWS
UDI-Device Identifier05414734507974
UDI-Public05414734507974
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P910023
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 11/25/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/31/2021
Device Model NumberCD1357-40Q
Device Catalogue NumberCD1357-40Q
Device Lot NumberA000088812
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/11/2020
Initial Date FDA Received11/25/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/31/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-