SMITH & NEPHEW, INC. GII NON POR FIN PUNCH 3-4; PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER
|
Back to Search Results |
|
Model Number 71440266 |
Device Problem
Dull, Blunt (2407)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/03/2020 |
Event Type
malfunction
|
Event Description
|
It was reported that, during set up, the device was dull.The procedure finished with the same device.No surgical delay.Patient injuries were not reported.
|
|
Manufacturer Narrative
|
The device, intended for use in treatment, was returned for evaluation.A visual inspection of the returned device confirmed the stated failure mode.The cutting edge of the device is dull and worn rendering it inoperable.The device shows signs of extensive use.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.A review of complaint history did not reveal additional complaints for the listed batch.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
|
|
Search Alerts/Recalls
|
|
|