WAVELIGHT GMBH ALLEGRETTO WAVE EYE-Q EXCIMER LASER; OPHTHALMIC EXCIMER LASER SYSTEM Back to Search Results
Catalog Number 8065990739
Device Problem Patient-Device Incompatibility (2682)
Patient Problems Blurred Vision (2137); Visual Disturbances (2140); Halo (2227)
Event Date 11/05/2020
Event Type  Injury  
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.No udi required due to this device being out of production prior to the september 24, 2014 udi regulation date.(b)(4).
 
Event Description
A healthcare professional reported a patient experienced halos, glare, shadows and fuzzy vision following lasik procedure.The patient notes hazy vision in the left eye.Additional information has been requested.
 
Manufacturer Narrative
A review of the device history record traceable to the reported serial number indicates that the product was processed and released according to the product¿s acceptance criteria.A review of the technical service onsite history showed no abnormalities that could have contributed to this event: laser was successfully verified prior to and after the day of treatment.The root cause could not be identified conclusively.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALLEGRETTO WAVE EYE-Q EXCIMER LASER
Type of Device
OPHTHALMIC EXCIMER LASER SYSTEM
Manufacturer (Section D)
WAVELIGHT GMBH
am wolfsmantel 5
erlangen 91058
GM  91058
MDR Report Key10911387
MDR Text Key218497669
Report Number3003288808-2020-00752
Device Sequence Number1
Product Code LZS
Combination Product (y/n)N
PMA/PMN Number
P020050
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 01/07/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue Number8065990739
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 11/05/2020
Initial Date FDA Received11/26/2020
Supplement Dates Manufacturer Received01/04/2021
Supplement Dates FDA Received01/07/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
-
-