|
Model Number CATRXKIT |
Device Problem
Failure to Advance (2524)
|
Patient Problems
No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/03/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
Results: the guidewire lumen was punctured approximately 6.0 cm from the distal tip.Conclusions: evaluation of the returned catrx revealed that the guidewire lumen was punctured.If the guidewire is forcefully advanced through the catrx guidewire lumen at an extreme angle, the guidewire may puncture the lumen.If the catrx is attempted to be advanced into a guide catheter with the guidewire outside of the guidewire lumen, resistance may be encountered, and the catrx may not be advance through the guide catheter.During functional testing, a demonstration guidewire was able to advance into the guidewire lumen without issue.Then the catrx with the guidewire inside the guidewire lumen was able to advance through a demonstration benchmark without issue.Penumbra catheters are inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|
|
Event Description
|
The patient was undergoing a thrombectomy procedure in the left anterior descending artery (lad) using an indigo system catrx aspiration catheter (catrx), a guide catheter, and a guidewire.During the procedure, the catrx was introduced approximately one third of the way into the guide catheter before it would not advance any further.Therefore, the catrx was removed.The procedure was completed using a new catrx and the same guide catheter.There was no report of an adverse effect to the patient.
|
|
Search Alerts/Recalls
|
|
|