One 777f8 with a monoject limited volume syringe and a non-edwards contamination shield were returned for examination.The reported event of pressure measurement issue was not confirmed during evaluation.The pressure monitoring kit was not returned with the catheter.All through lumens were patent without any leakage or occlusion.No visible damage or inconsistency was observed from the catheter body, syringe or balloon.The balloon inflated clear, concentric and remained inflated for 5 minutes without leakage.No fault messages showed up on the laboratory vigilance ii monitor when the catheter was connected.The thermistor and thermal filament circuit were continuous, there were no open or intermittent conditions.No visible inconsistence was observed on the eeprom data.The resistance value of thermal filament circuit was measured and found to be within specification.The reported event could not be confirmed or replicated during the analysis, as the device responded appropriately during functional testing.It is not known if some procedural factors may have contributed to the event.There was no evidence of a manufacturing nonconformance.No further actions will be taken at this time.A review of the manufacturing records indicated that the product met specifications upon release.Invasive procedures involve some patient risks.Although serious complications are relatively uncommon, the physician is advised, before deciding to insert or use the catheter, to consider the potential benefits in relation to the possible complications.The techniques for insertion, methods of using the catheter to obtain patient data information, and the occurrence of complications are well described in the literature.With any hemodynamic monitoring, pressure readings can change quickly and dramatically.Pressure tubing that is used with swan ganz catheters can also be a contributing factor to inaccurate values.In regards to the pressure tubing used with swan ganz catheters, it should be noted that poor dynamic response can be caused by air bubbles, clotting, excessive lengths of tubing, excessively compliant pressure tubing, small bore tubing, loose connections, or leaks.Pressure readings should correlate with the patient¿s clinical manifestations.It is unknown whether user or procedural factors contributed to the stated event.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis, and any excursions above the control limits are assessed and documented as part of this monthly review.Udi (b)(4).
|
It was reported that once the swan was placed in the patient, the pa pressures consistently stayed around 50/50.Troubleshooting included re-zeroing several times and unplugging and starting all over.When the stopcock was in the up position to zero out the pa line, the reading remained 50/50.In addition, no error messages were observed.The swan was removed and a new swan was obtained and placed into patient without any issues.No patient complications were reported.Patient demographics were unable to be obtained.
|