U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC IRELAND MICRA; INTRODUCER, CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC IRELAND MICRA; INTRODUCER, CATHETER Back to Search Results
Model Number MI2355A
Device Problem Use of Device Problem (1670)
Patient Problems Hemorrhage/Bleeding (1888); Low Blood Pressure/ Hypotension (1914); Hypovolemic Shock (1917); Loss of consciousness (2418)
Event Date 11/06/2020
Event Type  Injury  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that during implant of the leadless implantable pulse generator (ipg), the patient experienced decrease in consciousness and blood pressure, shock symptoms and bleeding of the abdominal cavity.Venography of the lower extremities revealed leakage of contrast medium from the right side at the left and right bifurcations.An emergency laparotomy was performed and hemostasis was performed at all bleeding points.The physician estimated that the puncture point was high, and the abdominal cavity and retroperitoneum were penetrated and a vein was hit.When the system was removed after the procedure, blood flowed into the abdominal cavity from the puncture site of the vein.It was diagnosed as decrease in consciousness level due to hemorrhagic shock.The leadless ipg remains in use.No further patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was further reported that there was no perforation inside the "true" lumen.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MICRA
Type of Device
INTRODUCER, CATHETER
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway,ie
EI 
MDR Report Key10929037
MDR Text Key218998807
Report Number9612164-2020-04714
Device Sequence Number1
Product Code DYB
Combination Product (y/n)N
PMA/PMN Number
K132030
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 12/18/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/18/2022
Device Model NumberMI2355A
Device Catalogue NumberMI2355A
Device Lot Number00151129
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 11/06/2020
Initial Date FDA Received12/01/2020
Supplement Dates Manufacturer Received12/18/2020
Supplement Dates FDA Received12/18/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Life Threatening; Required Intervention;
Patient Age71 YR
-
-