• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. OES CYSTONEPHROFIBERSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. OES CYSTONEPHROFIBERSCOPE Back to Search Results
Model Number CYF-5
Device Problems Device Reprocessing Problem (1091); Material Rupture (1546)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
The device referenced in this case has been received by olympus for physical evaluation.The investigation is ongoing.The definitive cause of the customer's experience can not be determined at this time.This report will be updated upon completion of the investigation or upon receipt of additional relevant information.
 
Event Description
It is reported during reprocessing of an oes cystonephrofiberscope, the scope was sterilized without the sterilization cap.As a result the bending cover ruptured.There was no patient contact/impact.
 
Manufacturer Narrative
This report is being updated to provided additional evaluation findings.The reported problem was found to be due to the device being sterilized without the sterilization cap.As a result the bending cover ruptured.This is a not reportable device malfunction and further investigation is not indicated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OES CYSTONEPHROFIBERSCOPE
Type of Device
CYSTONEPHROFIBERSCOPE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key10929531
MDR Text Key227214402
Report Number8010047-2020-09669
Device Sequence Number1
Product Code FAJ
UDI-Device Identifier04953170411144
UDI-Public04953170411144
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K032092
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 01/05/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCYF-5
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/12/2020
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/12/2020
Initial Date FDA Received12/01/2020
Supplement Dates Manufacturer Received12/02/2020
Supplement Dates FDA Received01/05/2021
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-