• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO CARDIOVASCULAR SYSTEMS CORPORATION ADVANCED PERFUSION SYSTEM 1; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS-ADVANCED PERFUSION SYSTEM 1

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CARDIOVASCULAR SYSTEMS CORPORATION ADVANCED PERFUSION SYSTEM 1; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS-ADVANCED PERFUSION SYSTEM 1 Back to Search Results
Model Number 801188
Device Problem Calibration Problem (2890)
Patient Problems No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/06/2020
Event Type  malfunction  
Manufacturer Narrative
Per the field service representative (fsr), the perfusionist had no issues with the gas flow out of the oxygen (o2) line into the oxygenator.The fsr found that the o2 sensor was not fully seated.He replaced the o2 sensor.Verification/release testing was performed successfully.The unit operated to the manufacturer's specifications.
 
Event Description
It was reported that during set-up of the device for a cardiopulmonary bypass (cpb) procedure, the electronic patient gas system (epgs) would not pass calibration multiple times.As mitigation, local control knobs were used to control the fraction of inspired oxygen (fio2).The surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.
 
Manufacturer Narrative
Updated block: d10.
 
Manufacturer Narrative
The reported complaint was confirmed.During laboratory analysis, the product surveillance technician (pst) placed the customer oxygen (o2) sensor into a lab use only electronic patient gas system (epgs).He observed the o2 sensor to pass repeated calibrations and track the fraction of inspired oxygen (fio2) within the expected tolerances per the release testing.Per data log analysis, on 06-nov-2020, each time calibration was performed it failed with an 'o2 sensor out of range at calib' message (o2 sensor value = 0).This indicates a bad sensor or a connection to the sensor.The sensor was replaced and calibration passed.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVANCED PERFUSION SYSTEM 1
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS-ADVANCED PERFUSION SYSTEM 1
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
6200 jackson road
ann arbor MI 48103
MDR Report Key10931609
MDR Text Key219523561
Report Number1828100-2020-00467
Device Sequence Number1
Product Code DTQ
UDI-Device Identifier00886799000588
UDI-Public(01)00886799000588(11)071211
Combination Product (y/n)N
PMA/PMN Number
K172220
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup,Followup
Report Date 02/26/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number801188
Device Catalogue Number801188
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/01/2020
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/06/2020
Initial Date FDA Received12/02/2020
Supplement Dates Manufacturer Received12/01/2020
02/22/2021
Supplement Dates FDA Received12/16/2020
02/26/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
OXYGENATOR.
-
-