|
Model Number LMD-2760MD |
Device Problem
No Display/Image (1183)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/02/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).
|
|
Event Description
|
It was reported that during the set up before the surgery the sony 27 monitor is powered on, the screen has lines across, as per this issue, the staff was unable to view images.No patient injuries or delay reported.Smith and nephew back-up device was available to complete the surgery.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
|
|
Manufacturer Narrative
|
H10: h2, h3, h6.The reported device, intended for use in treatment, was received for evaluation.There was a relationship found between the returned device and the reported incident.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review concluded this was an isolated event.A visual inspection was performed and no deficiencies were observed.A functional test revealed that the monitor powers up but displays red lines in place of an image.The complaint was verified and the root cause was associated with electrical component failure.
|
|
Search Alerts/Recalls
|
|
|