• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. EVIS LUCERA DUODENOVIDEOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. EVIS LUCERA DUODENOVIDEOSCOPE Back to Search Results
Model Number TJF-260V
Device Problem Device Contamination with Chemical or Other Material (2944)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
The subject device in this report has not been returned to omsc for evaluation.The exact cause of the reported event could not be conclusively determined at this time.If additional information becomes available, this report will be supplemented.
 
Event Description
Olympus medical systems corp.(omsc) was informed that it was found that the inside of the light guide lens was dirty during the incoming inspection of the subject device for the repair at the service department of olympus (b)(4).There was no report of patient injury associated with this event.
 
Manufacturer Narrative
This supplemental report is being submitted to provide additional information.The subject device was not returned to olympus medical systems corp.(omsc) for evaluation but returned to the service department of olympus south korea.Omsc confirmed the following information from the service department of olympus south korea.There was a brown part inside the light guide (lg) lens, and there were traces of fluid invasion.The fluid could have invaded through the surrounding of the lg lens, and omsc determined that there was a gap in the adhesive area around the lg lens.There were scratched on the lg lens which could have been attributed to physical stress.Omsc reviewed the manufacture history (dhr) of the subject device and confirmed no irregularity.The exact cause of the reported phenomenon could not be conclusively determined.However, based upon the past similar cases, there was the possibility that the reported phenomenon was attributed to the following causes.Physical stress.Chemical stress.Storage environment (direct sunlight, high temperature, high humidity, environment exposed to x-rays and ultraviolet rays, etc.).Ageing deterioration.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS LUCERA DUODENOVIDEOSCOPE
Type of Device
DUODENOVIDEOSCOPE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
MDR Report Key10933938
MDR Text Key219517466
Report Number8010047-2020-09716
Device Sequence Number1
Product Code FDT
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 01/14/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberTJF-260V
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/04/2020
Initial Date FDA Received12/02/2020
Supplement Dates Manufacturer Received01/05/2021
Supplement Dates FDA Received01/14/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-