• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOMET FRANCE S.A.R.L. OPTIPAC 80 REFOBACIN BONE CEMENT R-3; BONE CEMENT, ANTIBIOTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMET FRANCE S.A.R.L. OPTIPAC 80 REFOBACIN BONE CEMENT R-3; BONE CEMENT, ANTIBIOTIC Back to Search Results
Catalog Number 4712500398-3
Device Problem Improper Chemical Reaction (2952)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/17/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Report source, foreign - event occurred in (b)(6).The investigation is in progress.Once the investigation has been completed, a follow-up mdr will be submitted.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.
 
Event Description
It was reported that two optipac 80 were mixed in succession, in both cases the cement was difficult to mix.After 4 minutes, the cement had already hardened and had to be discarded.This mixing issue has led to a delay of 40 min during the surgery.No adverse event has been reported for the patient.
 
Manufacturer Narrative
This is a combination product (b)(4).This follow-up report is being submitted to relay additional information.No further information provided (x-rays, surgical report, photographs, lab test).The product analysis can't be performed as the product was not returned.The review of the device manufacturing quality can't be performed as the product batch number was not communicated.A complaint extract was done regarding cement paste too short setting time: - 1 complaint (2 products), this one included, has been recorded on optipac 80 refobacin bone cement r-3, reference (b)(4), from january 01, 2017 to february 10, 2021.- the complaint history, regarding the batch number, can't be performed as it was not communicated.According to available data, root cause of the event was unable to be determined.However, there is no evidence that the event is related to the product.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
It was reported that two optipac 80 were mixed in succession, in both cases the cement was difficult to mix.After 4 minutes, the cement had already hardened and had to be discarded.This mixing issue has led to a delay of 40 min during the surgery.No adverse event has been reported for the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OPTIPAC 80 REFOBACIN BONE CEMENT R-3
Type of Device
BONE CEMENT, ANTIBIOTIC
Manufacturer (Section D)
BIOMET FRANCE S.A.R.L.
plateau de lautagne bp75
valence cedex 26903
FR  26903
MDR Report Key10934456
MDR Text Key219545620
Report Number3006946279-2020-00231
Device Sequence Number1
Product Code MBB
Combination Product (y/n)N
PMA/PMN Number
N/A
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Type of Report Initial,Followup
Report Date 02/19/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number4712500398-3
Device Lot NumberNOT COMMUNICATED
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/18/2020
Initial Date FDA Received12/02/2020
Supplement Dates Manufacturer Received02/22/2021
Supplement Dates FDA Received02/22/2021
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A
Patient Sequence Number1
-
-