• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BLUE BELT TECHNOLOGIES REAL INTELLIGENCE ROBOTIC DRILL; ORTHOPEDIC STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BLUE BELT TECHNOLOGIES REAL INTELLIGENCE ROBOTIC DRILL; ORTHOPEDIC STEREOTAXIC INSTRUMENT Back to Search Results
Model Number ROB10013
Device Problem Suction Problem (2170)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/11/2020
Event Type  malfunction  
Event Description
It was reported that during a cori tka procedure, there was debris "flying" off the sterile field and bouncing back into the field from the drill/handpiece when milling bone.This is the fifteenth of seventeen incidents.No other complications were reported.
 
Manufacturer Narrative
G3, h2, h3, and h6: the product, ri robotic drill, rob10013, used for treatment was not returned for evaluation; thus, a visual and functional evaluation could not be performed, and a relationship between the reported event and the device could not be determined.While all products meet required manufacturing specifications prior to release a serial number or lot number is required to link the device to a dhr or nc investigation.A complaint history review for similar reported/confirmed complaints has identified prior events.Although the reported problem was not confirmed, factors that may have contributed to the reported symptom may be associated with improper assembly of tubing resulting in inadequate irrigation and suction, failure to clear surfaces of residual debris using irrigation, suction and/or hand tool orientation with respect to bone surface, limiting debris.Although no further containment or corrective action is recommended or required at this time, all complaints are monitored and trended through post market surveillance activities.If the product associated with this event is returned or provided at a future date, this evaluation will be reopened for investigation.D10: add concomitants.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REAL INTELLIGENCE ROBOTIC DRILL
Type of Device
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
MDR Report Key10935151
MDR Text Key219176677
Report Number3010266064-2020-02077
Device Sequence Number1
Product Code OLO
UDI-Device Identifier00885556757321
UDI-Public00885556757321
Combination Product (y/n)N
PMA/PMN Number
K191223
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 09/23/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberROB10013
Device Catalogue NumberROB10013
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 11/11/2020
Initial Date FDA Received12/02/2020
Supplement Dates Manufacturer Received09/23/2021
Supplement Dates FDA Received09/23/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
CORI ROBOTICS USA/(B)(6)
-
-