MAUDE Adverse Event Report: ABBOTT VASCULAR 20/30 PRIORITY PACK INDEFLATOR; ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR 20/30 PRIORITY PACK INDEFLATOR; ACCESSORIES Back to Search Results
Model Number 1000184
Device Problem Leak/Splash (1354)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/09/2020
Event Type  malfunction  
Manufacturer Narrative
The customer reported the device is not returning.Investigation is not yet complete.A follow up report will be submitted with all additional relevant information.
 
Event Description
It was reported that during use of a 20/30 indeflator there was difficulty inflating the unspecified device.There were no adverse patient effects reported and no clinically significant delay.No additional information was provided.
 
Manufacturer Narrative
The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other incidents from this lot.As the device was not returned for analysis, the investigation determined a conclusive cause for the reported difficulties cannot be determined.There is no indication of a product quality issue with respect to manufacture, design or labeling.
 
Manufacturer Narrative
Subsequent to filing the initial report, 48 unused/sterile devices were received.A random sample of 13 returned unused sterile devices were tested with no anomalies noted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
20/30 PRIORITY PACK INDEFLATOR
Type of Device
ACCESSORIES
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
MDR Report Key10935314
MDR Text Key219186760
Report Number2024168-2020-10093
Device Sequence Number1
Product Code MAV
UDI-Device Identifier08717648013591
UDI-Public08717648013591
Combination Product (y/n)N
PMA/PMN Number
K961471
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup,Followup
Report Date 10/29/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/31/2021
Device Model Number1000184
Device Catalogue Number1000184
Device Lot Number60198527
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 11/09/2020
Initial Date FDA Received12/02/2020
Supplement Dates Manufacturer Received12/07/2020
10/07/2021
Supplement Dates FDA Received12/30/2020
10/29/2021
Is This a Reprocessed and Reused Single-Use Device? No