• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. UNK NEXEL STEM; PROSTHESIS, ELBOW

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. UNK NEXEL STEM; PROSTHESIS, ELBOW Back to Search Results
Device Problems Adverse Event Without Identified Device or Use Problem (2993); Insufficient Information (3190)
Patient Problems Bone Fracture(s) (1870); No Information (3190)
Event Date 11/11/2020
Event Type  Injury  
Manufacturer Narrative
(b)(4).The device will not be returned for analysis; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.
 
Event Description
It was reported that a patient underwent an elbow arthroplasty on an unknown date.Subsequently, the patient is being considered for a revision on an unknown day for an unknown reason.However, no revision procedure has been reported to date.Attempts have been made and additional information on the reported event is unavailable at this time.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The device will not be returned for analysis; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.
 
Event Description
It was reported that the patient underwent a revision surgery due to periprosthetic fracture.The humeral component was revised and a longer stem was implanted.Attempts have been made and no further information is available.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated: b4, b5, g4, g7, h2, h6, h10 radiographs were provided and reviewed by a health care professional.Review of the available records identified the following: large periprosthetic lucencies related to loosening.Periprosthetic fracture also noted.Device history record review was unable to be performed as the lot number of the device involved in the event is unknown.A definitive root cause could not be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No further event information is available at the time of this report.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Additional information did not change the previous investigation.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No further event information is available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNK NEXEL STEM
Type of Device
PROSTHESIS, ELBOW
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
MDR Report Key10942363
MDR Text Key219491066
Report Number0001822565-2020-03929
Device Sequence Number1
Product Code KWI
Combination Product (y/n)N
PMA/PMN Number
NI
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup,Followup,Followup
Report Date 01/15/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/09/2020
Initial Date FDA Received12/03/2020
Supplement Dates Manufacturer Received12/16/2020
12/17/2020
01/15/2021
Supplement Dates FDA Received12/17/2020
12/30/2020
01/19/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Other; Required Intervention;
-
-