Note: reference 30400 is not registered in the united states.The u.S.Similar device is product reference 30400-01.A customer in (b)(6) notified biomerieux of out of range high results in association with the vidas(r) tsh 60 tests (ref.30400 lot 1007920110)when testing two external biorad quality control samples.Sample 1 biorad lyphochek immunoassay plus control ranges are as follows: the expected result was 0.420 uui/ml the expected bio-rad range = 0.203 - 0.637 uui/ml the expected rilibak range (13.5% deviation) = 0.363 - 0.476 uui/ml (b)(6) are the (b)(6) federal medical council guidelines similar to (b)(6).The customer tested sample one (1) six (6) times.One result was above the expected biorad range with a value of 0.82 uui/ml and all results were above the expected rilibak range.As there is no patient associated with this quality control strain, there is no adverse event related to any patient's state of health.
|