MAUDE Adverse Event Report: ASCENSION ORTHOPEDICS CADENCE IMPLANT-UNSPECIFIED; CADENCE TOTAL ANKLE SYSTEM|

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ASCENSION ORTHOPEDICS CADENCE IMPLANT-UNSPECIFIED; CADENCE TOTAL ANKLE SYSTEM| Back to Search Results
Catalog Number XXX-CADENCE IMPLANT
Device Problem Patient Device Interaction Problem (4001)
Patient Problem Pain (1994)
Event Type  Injury  
Manufacturer Narrative
Complaint sample was not returned for evaluation and lot number information has not been provided; therefore, an evaluation of the device could not be performed, and manufacturing records could not be reviewed.The cause(s) of the difficulty reported by the customer could not be determined.If additional relevant information becomes available in the future, this complaint will be reopened, and the respective evaluation performed.Trends will be monitored for this and similar issues.At present, we consider this complaint to be closed.According to risk management documentation for the cadence total ankle system, there are several identified failure modes which could result in implant impingement on hard/soft tissue, including fixation of the talar dome violates the subtalar joint; medial and/or lateral malleolus is damaged during bone resection; and incorrect talar trial size selected due to missing or incorrect laser markings.Per the risk management documentation, potential causes for tissue impingement include damage to tissue during bone preparation or insertion and implantation of incorrect sizes for the patient's anatomy.
 
Event Description
A facility reported a case from a study with cadence total ankle system: left ankle impingement and left medial malleolar stress reaction."the subject came into the clinic (b)(6) 2019 and was diagnosed with l.Ankle impingement.Subject later had a ct scan taken at an outside hospital and was diagnosed with l.Malleolar stress reaction.He elected to undergo surgical intervention after conservative methods of pt failed to relieve him of pain." surgical treatment was performed: "l.Ankle open debridement with partial excision of the medial malleolus and talar neck (osteophyte excision) with l.Ankle internal fixation of medial malleolus.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CADENCE IMPLANT-UNSPECIFIED
Type of Device
CADENCE TOTAL ANKLE SYSTEM|
Manufacturer (Section D)
ASCENSION ORTHOPEDICS
8700 cameron road #100
8700 cameron road #100
austin TX 78754
Manufacturer (Section G)
ASCENSION ORTHOPEDICS
8700 cameron road #100
austin TX 78754
Manufacturer Contact
vivian nelson
1000 campus drive
princeton, NJ 08536
6099362319
MDR Report Key10945991
MDR Text Key219516821
Report Number1651501-2020-00020
Device Sequence Number1
Product Code HSN
Combination Product (y/n)N
PMA/PMN Number
K151459
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other
Type of Report Initial
Report Date 11/09/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue NumberXXX-CADENCE IMPLANT
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 12/01/2020
Initial Date FDA Received12/04/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? Yes
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age56 YR
-
-