This supplemental report is being submitted to provide additional information.The device was not returned to olympus medical systems corp.(omsc), but returned to olympus repair center for the evaluation.According to the evaluation, olympus repair center found that the connector socket lock or connector pin was worn and broken.Device history record review indicates that the product was manufactured and tested in accordance with all applicable procedures and met all final product release criteria.The exact cause of the reported event could not be conclusively determined.Aging deterioration may have caused the defect because over 6 years have passed since the device was manufactured.If additional information becomes available, this report will be supplemented.
|