• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. HIGH FLOW INSUFFLATION UNIT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. HIGH FLOW INSUFFLATION UNIT Back to Search Results
Model Number UHI-3
Device Problem Pressure Problem (3012)
Patient Problem No Patient Involvement (2645)
Event Date 09/16/2020
Event Type  malfunction  
Manufacturer Narrative
The device was returned to the service center for evaluation.The evaluation found that there was a decrease in pressure from the unit.The investigation is still ongoing, and this report will be supplemented when new information becomes available.An olympus (b)(4) capa has been opened to manage the actions related to remediation of this late mdr reporting.
 
Event Description
The service center was informed that a leak was noted on the unit¿s regulator valve.There was no patient involvement reported.
 
Manufacturer Narrative
This supplemental report is being submitted to provide the additional information and device evaluation results.Please see the updates in sections: g4, g7, h2, h6 and h10.The legal manufacturer (lm) reviewed the content of this complaint for further investigation.The legal manufacturer reported that the root cause could not be determined.The lm reported that the most probable causes for the reported event are as follows: since more than 15 years have passed since the product was manufactured, it is presumed that it occurred due to wear caused by aging deterioration.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIGH FLOW INSUFFLATION UNIT
Type of Device
HIGH FLOW INSUFFLATION UNIT
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
MDR Report Key10951686
MDR Text Key226702054
Report Number8010047-2020-09869
Device Sequence Number1
Product Code HIF
UDI-Device Identifier04953170140280
UDI-Public04953170140280
Combination Product (y/n)N
PMA/PMN Number
K014166
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 01/15/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberUHI-3
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/16/2020
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 09/18/2020
Initial Date FDA Received12/05/2020
Supplement Dates Manufacturer Received12/18/2020
Supplement Dates FDA Received01/15/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-