• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. HIGH FLOW INSUFFLATION UNIT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. HIGH FLOW INSUFFLATION UNIT Back to Search Results
Model Number UHI-4
Device Problems Display Difficult to Read (1181); Electrical /Electronic Property Problem (1198)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/19/2020
Event Type  malfunction  
Manufacturer Narrative
The subject device in this report has not been returned to omsc for evaluation.The exact cause of the reported event could not be conclusively determined at this time.If additional information becomes available, this report will be supplemented.
 
Event Description
Olympus medical systems corp.(omsc) was informed that the set/actual flow rate indicator of the subject device could not be partially displayed.There was no report of patient injury associated with this event.The service department of olympus (b)(4) (oekg) checked the subject device and found that the reported phenomenon was duplicated.The user facility did not provide other detailed information.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's investigation and to correct information provided on the initial report.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the suggested event occurred because the light on the front panel was damaged, causing the set/actual flow rate indicator to be partially displayed.If additional information becomes available, a supplemental report will be executed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIGH FLOW INSUFFLATION UNIT
Type of Device
HIGH FLOW INSUFFLATION UNIT
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
MDR Report Key10955633
MDR Text Key226684781
Report Number8010047-2020-09896
Device Sequence Number1
Product Code HIF
Combination Product (y/n)N
PMA/PMN Number
K122180
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 01/29/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberUHI-4
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/24/2020
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/19/2020
Initial Date FDA Received12/05/2020
Supplement Dates Manufacturer Received01/21/2021
Supplement Dates FDA Received01/29/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-