• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. UNKN TAYLOR SPATIAL FRAME EXFIX DEV; PIN, FIXATION, THREADED, METALLIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. UNKN TAYLOR SPATIAL FRAME EXFIX DEV; PIN, FIXATION, THREADED, METALLIC Back to Search Results
Catalog Number UNKNOWN
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Non-union Bone Fracture (2369)
Event Type  Injury  
Manufacturer Narrative
(b)(4).
 
Event Description
"acta orthop traumatol turc 2016.Treatment of infected nonunion of the juxta-articular region of the distal tibia." author: göksel dikmen et al.In this study there were 13 patients, only 1 patient bearing tsf.It was reported that a patient suffered from persisting nonunion after tsf.Kirschner wire breakage occurred at the distal part of the fixator.The event was treated via revision surgery (retrograde intramedullary nail, the distal part of the fixator was replaced).
 
Manufacturer Narrative
The device, used for treatment, was not returned for evaluation.Therefore, product analysis could not be performed at this time.So the reported event could not be confirmed.A medical investigation was conducted and confirms the data presented in the aged article does not provide insight or relevance to current clinical outcomes for the product/device.Without clinically relevant patient-specific supporting documentation, a thorough medical investigation cannot be performed.The root cause and/or patient outcome beyond that which was documented in the article cannot be confirmed nor concluded; therefore, no further medical assessment is warranted at this time.Factors and/or some potential probable causes that could contribute to the reported event have been identified as overuse or excessive pressure on the joint, injury, infection and/or patient condition.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Without the actual product involved and/or device information, our investigation cannot proceed.If the device or new information is received in the future, this complaint can be re-opened.No further actions are being taken at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKN TAYLOR SPATIAL FRAME EXFIX DEV
Type of Device
PIN, FIXATION, THREADED, METALLIC
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10956059
MDR Text Key219850507
Report Number1020279-2020-07031
Device Sequence Number1
Product Code NDM
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 04/27/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue NumberUNKNOWN
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 11/11/2020
Initial Date FDA Received12/06/2020
Supplement Dates Manufacturer Received03/29/2021
Supplement Dates FDA Received04/27/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
-
-