C.R. BARD, INC. (BASD) -3006260740 VACCESS CT LOW-PROFILE POWER-INJECTABLE IMPLANTABLE PORT, CHRONOFLEX, 6F; PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR
|
Back to Search Results |
|
Model Number 7360000 |
Device Problems
Blocked Connection (2888); Loosening of Implant Not Related to Bone-Ingrowth (4002)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/09/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
As the lot number for the device was provided, a manufacturing review will be performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.(expiry date: 04/2021).
|
|
Event Description
|
It was reported that sometime post port placement, the hcp reports difficulty accessing the port and the port allegedly keeps moving in the pocket.It was further reported that the dye study was performed and the port condition was good.There was no reported patient injury.
|
|
Manufacturer Narrative
|
H10: manufacturing review: a lot history review was performed.This is the first complaint reported to date for this product and lot, therefore a device history record review is not required.Investigation summary: the device was not returned for evaluation.Therefore, the reported loosening of implant and blocked connection issues cannot be confirmed.A definitive root cause could not be determined.Labeling review: a review of product labeling documentation (e.G.Procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: d4 (expiry date: 04/2021).H11: section a through f ¿ the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : device not returned.
|
|
Event Description
|
It was reported that sometime post port placement, the nurse reports difficulty accessing the port and the port allegedly keeps moving in the pocket.It was further reported that the dye study was performed and the port condition was good.There was no reported patient injury.
|
|